Features 2
Bluelab Soil pH Meter 2
Introduction 3
Preparing for use 3
Calibration 4
Information about measuring
the pH of soils/media 6
Measuring the pH value 8
Measuring soil solution
pH value 8
Battery replacement 9
Cleaning the Bluelab
Soil pH Probe 9
Hydrating the soil pH probe 10
Storing the Bluelab Soil
pH Meter 10
Error messages 11
Technical speci cations 11
Troubleshooting guide 12
Bluelab Soil pH Probe
replacement 13
Bluelab Soil pH Probe Care 13
Product guarantee 14
Limitation of liability 14
Contact details 14
Care and use guide
Page view 0
1 2 3 4 5 6 ... 13 14

Summary of Contents

Page 1 - Care and use guide

soil pH meterContents pageFeatures 2Bluelab Soil pH Meter 2Introduction 3Preparing for use 3Calibration 4Information about measuring the pH of soil

Page 2 - Bluelab Soil pH Meter

English1010.0 Storing the Bluelab Soil pH MeterStore the soil pH meter in a cool, dry and clean place when not in use.Keep out of direct sunlight.Kee

Page 3 - 2.0 Preparing for use

English1111.0 Error messagesAn error message will only appear following pH calibration failure.‘Err’ will be displayed for a few seconds then the disp

Page 4 - 3.0 Calibration

English1213.0 Troubleshooting guideTrouble Reason CorrectionpH reading inaccurateContaminated soil pH probe / glassware not clean.Clean soil pH probe

Page 5 - 3.0 Calibration cont

English13Bluelab Probe Care - pHThe instrument is only as accurate as the probe is clean!Bluelab pH Probe KCl Storage SolutionThe perfect solution to

Page 6

English14Bluelab Soil pH Meter product guaranteeBluelab Corporation Limited guarantees this product for a period of5 years (60 months) from the date o

Page 7

English2Bluelab Soil pH MeterLow battery indicatorSuccessful pH calibration indicatorspH unitATTENTIONIf it dries, it dies!Keep your soil pH probe tip

Page 8 - 5.0 Measuring the pH value

English31.0 IntroductionThe Bluelab Soil pH Meter has two press buttons; ‘calibrate’ and power. The power button requires a short press; release in ab

Page 9 - 7.0 Battery replacement

English43.0 CalibrationpH calibration is required before fi rst use and then monthly to ensure readings are accurate.For accurate pH readings the so

Page 10 - Storage Solution

English53.0 Calibration cont.To calibrate the pHClean soil pH probe tip.See section 8.0 (the soil pH probe does not require cleaning before the fi r

Page 11 - 12.0 Technical speci cations

English64.0 Information about measuring the pH of soils/mediapH is the measurement of the hydrogen ion concentration (H+) - acidity and its opposit

Page 12 - 13.0 Troubleshooting guide

English74.0 Information about measuring the pH of soils/media cont.Taking pH measurements of soils with an electronic meter is indicative rather th

Page 13 - Solution

English85.0 Measuring the pH valueOnce the Bluelab Soil pH Meter has been set up and calibrated, using it to measure a pH value involves using the Bl

Page 14 - Limitation of Liability

English97.0 Battery replacementBatteries are replaced in the Bluelab Soil pH Meter when the low battery indicator appears on screen. The low battery i

Comments to this Manuals

No comments